NIRS™ DS3 for sugar cane processing

The innovative NIRS™ DS3 combines unique analytical performance with the latest software and networking services, making reliable test data for decision-making more available than ever

NIRS™ DS3 for sugar cane processing

Overview

Value

Save time and money in your evaluation of raw materials and bagasse production. Produce closer to specification and ensure consistent final product by controlling quality throughout your manufacturing process.

Sample Type

Shredded/prepared cane, cane juice, filter cake, syrup, massecuite, seed magma, crystal sugar, molasses and bagasse

Parameters

Pol, brix, purity, moisture, ash, RS, and more

Features

Automated sample transfer
Automated sample transfer

Automated sample transfer reduces time spent on transferring data to and from management systems

Smart Start
Smart Start

The Smart Start function ensures that your NIRS DS3 is ready to measure without having to wait for diagnostics. Just start the shift and start measuring.

FossManager™
FossManager™

FossManager™ software allows instruments to be connected in a network making it easier, quicker and more reliable to manage individual or multiple instruments from a desktop located anywhere.

Ready-to-use calibrations or tools to develop your own
Ready-to-use calibrations or tools to develop your own

FOSS global ANN calibrations are built on thousands of samples from a large variety of raw materials. Alternatively, develop your own calibrations with the easy to use FossCalibrator™ tool.

Versatile performance
Versatile performance

Advanced and versatile solution for testing a range of parameters on ground or liquid material with no or very limited sample preparation.

Technical Specifications

Products measured and product presentation
Feature Specification
Dimensions (w x d x h) 375 x 490 x 300 mm
Weight 27 kg
Degree of protection IP 65
Measurement mode Reflectance or transflectance (for liquids)
Wavelength range 400 – 2500 nm
Detector Silicon (400 – 1100 nm), Lead Sulphide, PbS (1100 – 2500 nm)
Optical bandwidth 8.75 ± 0.1 nm
Data resolution 0.5 nm
Number of data points 4200
Number of subsamples Default: 7 for small cup; 8 for large cup
Self test 12 minutes (variable)
Analysis time <1 minute for 32 scans (8 subsamples, 4 scans per subsample)
Network connections FossManager™
Installation requirements
Voltage supply 100-240 V AC*, frequency 50-60 Hz, Class 1, protective earth
Ambient temperature 5 – 40°C
Storage temperature -20 to 70°C
Ambient humidity < 93% RH
Mechanical environment Stationary during use
EMC environment Laboratory use, industry requirements
*Mains supply voltage fluctuations not exceeding ±10% of the rated voltage.