Infratec™

Solid, straightforward and reliable, Infratec™ draws on the latest advances in NIR technology, connectivity and usability.

It makes the job of quality control easier and less time-consuming as a reliable cornerstone for any grain handling operation.

Infratec™

Overview

Value

Makes your job easier with enhanced usability. Built on advanced NIR technology for the most trustworthy results. Improves uptime with digital connectivity and support package.

Sample Type

Barley, wheat, corn and other cereals, oilseeds, beans and pulses.

Parameters

Moisture, protein, oil, test weight, starch, wet gluten, fibre, ash and many more.

Features

ANN - a global industry needs global calibrations
ANN – a global industry needs global calibrations

The huge Infratec™ database comprises over 50.000 cross checked samples, PLS and robust ANN-based calibrations building on a wide sample range from many years of harvests. This gives a level of accuracy and stability that enables Infratec™ to test even the most unusual samples anywhere in the world.

FossConnect™
FossConnect™

FossConnect™ is a centralised networking management solution for Infratec™ NOVA that helps you reduce your production costs and secure the quality of your products. With FossConnect™, you can manage, configure and monitor your instruments remotely from any internet-enabled computer and safeguard your data for whenever you need it.

Established standard worldwide
Established standard worldwide

Infratec™ is officially approved and established worldwide as the standard for determining protein, moisture, oil and starch. Its accurate and reliable results help you to deliver consistent quality. With Infratec’s patented stabilising technique, you get correct results whatever the weather. For instance, protein is predictable to within 0,1% and it can be measured in temperatures ranging from -4°C to +40 °C.

Technical Specifications

Feature Specification
Dimensions (W x D x H) 410 x 460 x 415 mm
Weight 28.5 kg (31 kg with test weight module)
Voltage 220-240V 50-60Hz or 110-120V
Rated current 1.0A (110-120V) / 0.5A (220-240V)
Fuse T 5 A (250V)
Power consumption 85 W (24V)
Input supply 24V DC from FOSS approved power supply
Spectrometer Scanning monochromator
Wavelength range 570 – 1100 nm
Detector Silicon
Optimal bandwidth 7 nm
Number of data points/scan 1061
Mode Transmittance
Light source Tungsten halogen lamp
Interface Ethernet, 3 x USB (full function) including one on the instrument front for easy access
Display 10 inch capacitive touch screen
Noise Level < 70 dB(A)
Degree of protection IP 54
Sample handling and result presentation
Analysis time

Less than 60 seconds for 10 sub-samples including test weight analysis and as little as 40 seconds when dynamic sub-sampling enabled

Path length

Variable cell automatically controlled from 6 – 33 mm

Result report

Presented on the display as default. Can be sent to PC/LIMS and the printer port

Outlier function

Warnings and options for the presentation of the result

Software

Menu driven with touch screen interface

Regression programs

ANN (Artificial Neural Network); PLS (Partial Least Squares)

No. of sub-samples

Between 1 and 30 sub-samples (10 sub-samples standard)

Other Products by FOSS

Fibertec™ 1023

The Fibertec™ 1023 is designed for rapid determination of dietary fibre according to established enzymatic methods, including those approved by AOAC, AACC, NMKL and Asp. Includes a filtration module and a shaking water bath for quantitative determination of dietary fibre.