Overview
Value
First fully automated Soxhlet system. Improving throughput at a lower cost. Reducing risk in fat analysis. Batch handling tools for easy operation.
Sample Type
Feed, Feed ingredients, Forage, Pet food, Food, Food ingredients, Grain, Cereal, Oil seeds, DDGS, Meat, Meat products, Fish, Fish products, Dairy products, Environmental products, (e.g. Soil, Sludge, Waste water) and other industry products (Paper pulp, Textiles, Rubber, Tobacco, Fertilizer, Detergents, Plastic and more)
Parameters
Crude Fat, Total Fat and Extractable Matter
Features
Secure solvent dispensing & recovery
Flexible operation
The Soxtec™ 8000 Total Fat solution helps you respond to the demands of your customers, for example, you can do Total or Crude Fat analysis or both. It’s easy to increase capacity according to customer requirements with up to 12 position hydrolysis and six position extraction that is expandable to 12, all with individual hotplate control.
Simplified process
This unique patented Hydrocap™ filter contains the sample from the initial weighing, through hydrolysis to final extraction; avoiding sample transfer errors. It is a fully automated system that performs automated acid hydrolysis and Randall improved Soxhlet extraction in one integrated process.
Case Story
Models
Soxtec™ 8000 | ST 255 Soxtec™ | ST 243 Soxtec™ | |
Levels of automation | Fully automated | Semi automated | Manual |
Positions | 6 – 12 | 6 | 6 |
Capacity/day | 42/86 samples | 30 – 36 samples | 30 – 36 samples |
Solvent volume | 40 – 110 ml | 70 – 90 ml | 40 – 50 ml |
Thimble diameter | 26, 33 mm | 26, 33 mm | 26 mm |
Compatible with hydrolysis unit | Hydrotec™ 8000 and SC 247 SoxCap™ | Hydrotec™ 8000 and SC 247 SoxCap™ | SC 247 SoxCap™ with mini capsules |
Specification
Extraction time | Typically 45 – 60 min | |
Solvent volume | 40 – 110 ml depending on cup type | |
Sample size | 0.5 – 3 g depending on sample type 0.5 – 2 g (for total fat) |
|
Capacity/batch | 6/12 positions | |
Capacity/day | 42/84 samples | |
Temperature range | 0 – 285 °C | |
Measuring range |
0.1 – 100 % |
|
Reproducibility | 1% rel. Or better (5 – 100% fat) | |
Solvent Recovery |
Typically 80 % |
|
Accuracy | According to officially approved methods | |
Installation requirements | ||
Power supply |
200 – 240 V 50 – 60 Hz |
|
Power consumption |
1500 W (120 VAC version) |
|
Water consumption |
2 l/min < 25°C minimum flow |
|
Weight net |
Extraction Unit
35 kg |
Control unit
4 kg |
Dimensions (W×D×H) |
640 × 350 × 630 mm | 280 × 200 × 190 mm |