Overview
Value
High throughput with high uptime, the MilkoScan™ 7 is the ultimate tool for cost effective laboratory operations with flexible licensing options for a no-limits test portfolio.
Sample Type
Raw milk (cow, goat, sheep, buffalo).
Parameters
Fat, protein (true and crude), casein, lactose, solids (SnF and TS), urea, citric acid, fatty acid profiling, Free Fatty Acids (FFA), freezing point depression, pH, ketosis screening (BHB and acetone), untargeted raw milk screening (adulteration).
Features
Diamond cuvette for improved robustness ensures maximum uptime
MilkoScan™ 7 RM has a diamond cuvette backed by 10 years guarantee. Optics are optimized on the interferometer module increasing the signal/noise ratio, ensuring the same high performance at all cadences and improving repeatability for minor constituents.
Adulteration – untargeted raw milk screening
Raw milk containing abnormalities is a growing problem. The abnormalities can be caused by deliberate adulteration, for example, with lard or melamine or by accidents if water or cleaning agents are mixed with the milk. With the MilkoScan™ 7 RM you can screen incoming raw milk samples to identify a suspect raw milk sample quickly and as a normal part of everyday testing. The suspect sample can then be further analysed to determine the contaminant.
FossConnect™
FossConnect™ is a centralised networking management solution for MilkoScan™,Fossomatic™, BactoScan™, and CombiFoss™ that helps you to reduce operational costs and secure consistent high performance from your analytical instruments. With FossConnect™, you can manage, configure and monitor your instruments remotely from any internet-enabled computer and safeguard your data for whenever you need it.
FossAssure™
FossAssure™ is a MilkoScan™ digital service designed to reduce product waste, secure brand protection and increase profitability by ensuring the quality and precision of your analytical results. Besides, FossAssure™ services will also improve instrument uptime and reduce operational costs by optimising the operation and the performance of your instruments.
Technical Specifications
Specifications MilkoScan™ 7 RM | |||||
Most of the calibrations are using multiple wavelengths selected freely from the entire Mid-IR spectrum in order to optimize robustness and accuracy. Compared to traditional filter calibrations, they are called full spectrum calibrations. | |||||
Performance Carry-over for all components <1% relative |
|||||
Component |
Measuring range | Performance range | Repeatability | Accuracy Bulk | Accuracy Single cow |
Fat | 0-15% | 2-15% | Cv < 0.5% | Cv < 1.0% | Cv < 1.5% |
Protein | 0-10% | 2-10% | Cv < 0.5% | Cv < 0.9% | Cv < 1.5% |
Lactose | 0-10% | 2-10% | Cv < 0.5% | Cv < 0.9% | Cv < 1.5% |
Solids | 0-20% | 2-20% | Cv < 0.5% | Cv < 1.0% | Cv < 1.5% |
Urea | 10-100 mg/dl | 10-100 mg/dl | Sd < 1.5 mg/dl | Sd < 3 mg/dl | Sd < 3.5 mg/dl |
Citric Acid | 0.1-0.5% | 0.1-0.5% | Sd < 0.005% | Sd < 0.01% | Sd < 0.015% |
FPD (Screening) | 400-600 m°C | 450-550 m°C | Sd < 0.5 m°C | Sd < 4 m°C | N/A |
Novel parameters | |||||
Fatty acids profile | |||||
Ketosis screening (BHB, acetone) | |||||
Untargetted screening raw milk (adulteration) | |||||
Application data | |||||
Analysis Capacity | 100, 200, 300, 400, 500 or 600 samples per hour | ||||
Sample intake | 5 ml | ||||
Required sample temperature | 37 – 42°C | ||||
Performance Specifications | Full spectrum calibrations | ||||
Standards and Approvals MilkoScan™ 7 RM is CE-labelled and complies with the following directives and regulations:• EMC (ElectroMagnetic Compatibility) Directive 2004/108/EC • LVD (Low Voltage) Directive 2006/95/EC • Machinery Safety Directive 2006/42/EC • Regulation (EC) 1272/2008 on classification, labelling and packaging of substances and mixture, CLP (EC) • WEEE Directive 2002/96/EC • Packaging and packaging waste Directive 94/62/EC • REACH 1907/2006/ECThe MilkoScan™ 7 RM techniques comply with: • ISO 9622 / IDF 141:2013 • AOAC official method 972.16By using wavelengths from the entire Mid- IR spectrum for each component, calibrations are optimised in terms of robustness and/or accuracy (temperature, homogenisation and humidity) |