BacSomatic™

Know the hygiene quality of raw milk before you start using it with fast Bacteria and Somatic Cell Count in one handy instrument.

BacSomatic™

Overview

Value

Small and handy instrument for Bacteria Count and Somatic Cell Count offers a rapid alternative to manual assay or semi-automated methods requiring reagent handling. Reagent bag system ensures exact reagent dosage every time reducing the risk of human error that can occur with other methods.

Sample Type

Raw cow’s milk

Parameters

Bacteria count and somatic cell count.

Features

Rapid results for somatic cell count and bacteria count
Rapid results for somatic cell count and bacteria count

Results for bacteria count and somatic cell are delivered within nine and a half minutes (1.5 minutes for somatic cell alone). The method is the same as the well-known methods used in FOSS BactoScan™ and Fossomatic™  analysers used by raw milk testing centres around the world.

Flow system
Flow system

Unique flow system directs the sample to a bacteria incubator or somatic cell mixing chamber where the cells are stained and then measured by the laser source. Two different laser intensity levels are used according to whether a somatic cell count or bacteria count is to be performed.

Automated reagent dosage and handling
Automated reagent dosage and handling

Automated procedure ensures exact reagent dosage every time and avoids the need for reagent mixing. The reagent bag system is specifically designed for safe, contact-free use of reagents. The reagents are provided ready-to-use in bags. Place the bags in the instrument and close the door and you are ready to test.

Technical Specifications

Analysis capacity SCC and IBC :15 samples/hour
Only IBC :15 samples/hour
Only SCC: 40 samples/hour
Carry over IBC: <1% relative, usually <0.5%
SCC: <1% relative, usually <0.5%
Analysis time

9.5 minutes IBC
1.5 minutes SCC alone

Sample intake SCC and IBC: 6ml
Only IBS: 5.2 ml
Only SCC: 2.2 ml
Sample temperature 2 – 4°C
Network connections Mosaic
Dimensions (H*W*D*) 400 x 400 x 400 mm
Weight 25 kg
Working factor IBC : 100
SCC: 100
Measuring range IBC: 5000-20 mln IBC/ml
SCC: 0-10 mln cells/m
Performance range IBC: 10000-10 mln IBC/ml
SCC: 100000-1.5 mln cells/ml
Repeatability IBC:
Sr≤0.07 log relative, usually 0.06 at10-50 IBC/μl
Sr≤0.05 log relative, usually 0.04 at51-200 IBC/μl
Sr≤0.04 log relative, usually 0.02 at >200 IBC/μlSCC:
CV<6% at 100000 cells/ml
CV<4% at 300000 cells/ml
CV<3% at 500000 cells/ml
Accuracy IBC: Typical Sy,x <0.25 log units from SPC (plate counting)
SCC: <10% relative mean different from direct microscopic SCC